תיאור

Medical Face Mask Synthetic Blood Penetration Tester RH-CT30

RH-CT30 Mask synthetic blood penetration tester is used for the determination of penetration performance of synthetic blood and other liquid-derived liquid injection to evaluate its safety and protection performance.It is suitable for the determination of penetration of synthetic blood spatter in surgical mask.It is widely used in medical inspection department, safety inspection department and scientific research unit

 

Medical Face Mask Synthetic Blood Penetration Tester RH-CT30

הxecutive standard:

YY 0469-2011:surgical mask

GB19083-2010:Technical requirements for medical protective masks;

YY/T0691-2008:Method of test for the penetration of synthetic blood (fixed volume, horizontal injection);

 

Medical Face Mask Synthetic Blood Penetration Tester RH-CT30

טטכני פערכות:

Test environmentטֶמפֶּרָטוּרָה (21±5)℃;לחות יחסית (85±10)%
Injection distance(300±10)מ"מ
Nozzle diameterפי 0.84 מ"מ, 12.7 mm long
Liquid injection speed450cm/s;550 cm/s;635 cm/s
Spray pressure standard 10.7kpa 16kpa 21.3kpa/adjustable
ספק כוחAC (110~240) V 50Hz

 

Medical Face Mask Synthetic Blood Penetration Tester RH-CT30

פתכונת roduct:

Air pressure range can be set freely, set range (0.5– 30)kPa;

Touch screen Chinese display, friendly man-machine interface, aluminum alloy panel, equipped with metal buttons, ensure the instrument durable

3.The sample table is made of 304 stainless steel plate, which is light, clean and rust-free.

4.Constant pressure injection device, no manual adjustment

Medical Face Mask Synthetic Blood Penetration Tester RH-CT30

טest Method:

GB19083-2010Technical requirements for medical protective masks

Synthetic blood penetration test;

Spray 2 mL of synthetic blood into the mask at a pressure of 10.7 kPa (80 mmHg). No infiltration should occur inside the mask.

YY 0469-2011 Synthetic blood penetration test of surgical mask

Spray 2 ML of synthetic blood into the mask at a pressure of 16.0 kPa (120 mmHg). No infiltration should occur inside the mask

YY 0469-2011Synthetic blood penetration test of surgical mask:

סample size: Three samples were used for the experiment.

Sample pretreatment: the sample was pretreated at least 4 h in the environment of temperature (21 × 5)C and relative humidity (85 × 5)%, and the test was conducted within 1 min after taking out.

טest procedure:The sample is fixed on the sample fixture on the instrument, וה 2 mL table is placed 30.5cm away from the sample center

The surface tension of the synthetic blood was 0.002n /m at a pressure of 16.0 kPa (120 mmHg)

The needle tube with the diameter of 0.84mm was sprayed horizontally to the target area of the sample under test. Visual inspection was conducted within 10 s after removal.

 

Medical Face Mask Synthetic Blood Penetration Tester RH-CT30

רesult רeduction:

Check the inner side of the sample for penetration.If visual inspection is suspicious, use a absorbent cotton swab or similar to rub the inside of the target area and determine if there is synthetic blood penetration.The resulting structure shall meet the standard requirements.

 






    התווסף לעגלת הקניות שלך:
    לבדוק