Description

UL94 Horizontal vertical flammability test instrument is applicable to test and assess the burning characteristics of plastic material ,can do fire flammability test,vertical flammability test, and equip with gas flowmeter, adjust the flame size, simple and safe operation.This instrument can use to assess the flammability of materials or foam plastics such as V- 0, v-1, v-2, HB, 5V, HF- 1, HF- 2, HBF.

UL94 Horizontal vertical flammability test instrumen Test standard

fire test UL94

GBT2408-2008: determination of plastic combustion performance and vertical method

IEC 6065-11-10 fire prevention test

UL94 Horizontal vertical flammability test instrument Technical parameters:

Contents:≥ 0.5 cubic meter, glass observation door

Timer: the imported timer is adjustable within the range of 0 ~ 99 minutes, with a precision of plus or minus 0.1 seconds. The combustion time can be set and the combustion duration can record

Casting time: 0 ~ 99 minutes 99 seconds can  set

Residual flame time: 0 ~ 99 minutes 99 seconds can set

Afterburn time: 0 ~ 99 minutes 99 seconds can  set

Test gas: more than 98% methane /37MJ/m3 natural gas (gas also available)

Burning Angle: 20 °, 45 °, 90 ° (0 °) is adjustable

Size parameters of the burners are: import lamp itself, nozzle Ø 9.5±0.3 mm, inner diameter nozzle effective length to 100 ± 10 mm, have air conditioning

Flame height: the standard requirement adjust from 20mm to 175mm

Flow meter: adopt 105ml/min standard

Product features: other configurations include lighting fixtures, exhaust gas installations, gas flow regulating valves, gas pressure gauges, gas pressure regulating valves, gas flow meters, gas u-type pressure gauges and sample fixtures

Power supply: AC 220V, 50Hz

Character

1, the burner electric control adjustment position, automatic timing positioning, easy to operate.

2, equipped with highly sensitive combustible gas leakage alarm device, gas leakage automatic alarm;

3. The burner can be inching up or down, and the distance between the burner and the sample can be adjusted at will;

4, equipped with a special exhaust system, automatic exhaust gas;

5. There are two sets of sample clips for horizontal and vertical tests;

6. Switch for horizontal and vertical test to facilitate the conversion operation of horizontal test and vertical test.

Fixture

Photos for PLC controller

UL94 Horizontal vertical flammability test instrument More Photos

 






    has been added to your cart: