معايرة آلة التوتر بالكمبيوتر, معايرة آلة التوتر برامج الكمبيوتر, معايرة آلة اختبار التوتر

معايرة آلة التوتر بالكمبيوتر, برامج الكمبيوتر معايرة آلة التوتر, جهاز اختبار التوتر، واجهة معايرة مستشعر الطاقة, مستشعر السفر, rubber extenometer, metal extenometer and system Settings five TAB, select each TAB for the corresponding sensor calibration or setting.

1. Calibrate the force sensor

Select the force sensor TAB, كما هو مبين في الشكل 6.There are four force sensor channels shown on the left: “50كلغ”, “0.0كلغ”, “0.0كلغ” و “0.0كلغ”.

The calibration process is as follows:

1. Plug in the force sensor to be calibrated, such as the force sensor with a range of 50kg, and select a force sensor channel on the left;

2. Input the range of the sensor in the “يتراوح” input box, مثل “50” كلغ;

3. Input the zero-value force of the sensor in the zero-value input box, and generally input “0”;

4. Click theEmptybutton to empty the calibration table;

5. Calibration point 1: empty the sensor, return the force to zero, and input “0” kg in the input box to the right of thestandard weight value”. Then click theTake pointbutton, then the AD value and the standard weight value of point 1 will be taken into the calibration table;

6. Calibration point 2: Load the sensor with a certain value of load, such as hanging 5kg weights, input the load value of “5” kg in the input box on the right ofstandard weight value”, and click thetake pointbutton, then the AD value and standard weight value of point 2 will be taken into the calibration table;

7. Calibration point 3: The method is the same as point 2. If the calibration point 3 is continued, ال “AD value” و “standard weight valuecan only be larger than point 2, otherwise the calibration will be inaccurate;

8. Calibration point 4: The method is the same as point 2. If the calibration point 4 is continued, ال “AD value” و “standard weight valuecan only be larger than point 3, otherwise the calibration will be inaccurate;

9. Calibrate point 5: The method is the same as point 2. If the fifth point is continued to be calibrated, ال “AD value” و “standard weight valuecan only be larger than point 4, otherwise the calibration will be inaccurate;

10. Calibration point 6: The method is the same as point 2. If the calibration point 6 is continued, ال “AD value” و “standard weight valuecan only be larger than point 5, otherwise the calibration will be inaccurate;

11. Click the buttonWrite calibration data to acquisition Card”. Ifwrite successfullyis prompted, then the calibration of a force sensor will be completed.

12. If the second force sensor is needed, repeat steps 1-11 فوق; خلاف ذلك, set the range of all other sensors to 0, and then click the buttonWrite calibration data to acquisition Card”.

ملحوظة: على الأقل 2 points should be calibrated. Steps 7~10 May not be performed.According to the characteristics of most sensors in China at present, a 3-point calibration can generally achieve very good accuracy. The second point generally values 3-5% of the sensor range, and the third point generally values 30-50% of the sensor range.If more than 3 نقاط (مشتمل 3 نقاط) are calibrated, ال “AD value” و “standard weight valueof the latter point must be greater than that of the previous point, otherwise the calibration will be inaccurate.If the AD value decreases as the standard weight increases, it can be increased by switching the two signal lines of the force sensor.When there is no load, ال “AD valueis generally around 8400,000. When the sensor is loaded, ال “AD valuechanges in proportion. If theAD valueis unchanged when the sensor is loaded, it indicates that the hardware is not working normally, which may be due to the damage of the force sensor or the failure of the acquisition card.

If found after calibration measuring force value has small difference with standard weight, can be addressed by calibrated once again, also can be directly modified by calibration standard weight values in the table data, such as software measurement is bigger than standard weight value, the calibration can be in the table 2 points of standard weight value in a bit, then clickto write the calibration data into the acquisition cardbutton, reset again, until the two values are equal.

2. Calibrate the travel sensor

معايرة آلة التوتر بالكمبيوتر, معايرة آلة التوتر برامج الكمبيوتر, TAB of calibration and selection of travel sensor of tension tester, كما هو مبين في الشكل 7:

شكل 7.

1. The input box on the right side of thejourneycoefficient input coefficient value, travel coefficient refers to the acquisition card each received 1000 corresponding travel value, pulse acquisition card to 4 times the frequency of encoder line number, such as encoder line number is n, screw lead of Lmm corresponding coefficient of schedule for the L/n / 4 * 1000, if not clear encoder line number or screw lead, can be arbitrary input a testing values such as “1”;

2. Click the buttonWrite calibration data into the Acquisition card”, and write the value ofTravel factorinto the acquisition card;

3. Click theTrip to Zerobutton to reset thecurrent tripto zero;

4. Click the “أعلى” أو “downbutton to walk a certain distance of the machine, and measure the distance with the vernier caliper;

5. Check whether the value ofcurrent travelis equal to the measured value. If the value ofcurrent travelis larger than the measured value, please change the value ofTravel coefficientto a smaller value; خلاف ذلك, change the value to a larger one, and then click the buttonWrite calibration data to the Acquisition Cardto write the new value ofTravel coefficientto the acquisition card;

6. كرر الخطوات 3-5 until thecurrent strokevalue equals the measured value;

7. Complete the calibration of the travel sensor;

3. Calibrate the rubber extenometer

If the machine is connected with a rubber extensometer (also known as large deformation), the rubber extensometer needs to be calibrated. Before calibration, please be sure to setuser parameters” -& GT;Control parameters” -& GT;ال “الإزاحة” under “المستشعر” is selected asrubber extenometer”, otherwise the calibration of rubber extenometer cannot be carried out.The calibration interface of rubber extensometer is shown in Figure 8:

شكل 8.

معايرة آلة التوتر بالكمبيوتر, معايرة آلة التوتر برامج الكمبيوتر, calibration of rubber extenometer calibration process of tension testing machine with travel sensor.

4. Calibrate the metal extender

If the machine is connected with a metal extender, it is necessary to calibrate the metal extender. Before calibration, please make sure to setuser parameters” -& GT;Control parameters” -& GT;ال “الإزاحة” under “المستشعر” is selected asmetal extensometer”, otherwise the calibration of metal extensometer cannot be carried out.The calibration interface of metal extensometer is shown in FIG. 9:

شكل 9.

The calibration process of the metal extensometer is the same as that of the force sensor.