  1. Fundamentals

The pressure difference method is used to place the pre-processed sample between the upper and lower measuring surfaces, and a constant pressure difference is formed on both sides of the sample.Under the action of pressure difference, the gas flows from the high pressure side through the sample to the low pressure side, and the permeability of the sample is calculated according to the area, pressure difference and flow through the sample.

  1. Definition of breathability

Average air flow through paper and board per unit area, expressed in microns per pascal-second [μm/ (باس)], at unit time and unit pressure difference under specified conditions.Its calculation formula is as follows;Ps = V/(delta p, t)

أين: Ps — نفاذية, μm/Pa·s

الخامس — Volume of air passing through the sample during measurement time, مل

△ PPressure difference on both sides of the sample, كيلو باسكال

ت — Determine the time, س

  1. Unit of breathability

The unit of paper permeability isμm/ (باس) (micron per PASCAL second)”.1μm/ (باس) =1mL/ (m2·Pa·s) =1L/ (m2·kPa·s).على سبيل المثال, a paper with a breathability of 10μm/(باس) can be interpreted as 10mL/s (10mL/s) of air flowing through 1 square meter (m2) of paper at a pressure difference of 1 بنسلفانيا (بنسلفانيا).Or at a pressure difference of 1 كيلو باسكال, air flow through 1 square meter (m2) of paper is 10 liters per second (10لتر / ثانية).

في التمرين, breathability is also expressed in other units, such as CU [cm3/(min·cm2)] for cigarette paper, which refers to the air flow rate (ml /min cm3/min) across 1 square centimeter (سم2) of the surface of the tested sample at a pressure of 1 kPa.In theGellet methodbreathability meter, the breathability unit is also expressed asGellet second (س)”, which refers to the time taken to pass 100 mL gas in an average pressure difference of 1.23 kPa and test area of 6.42 مم2.

  1. Test area and pressure difference used by breathability tester

The test area and differential pressure of the laboratory breathability meter are specified in the standard.Common areas are: 10سم2, 6.42سم2, 2cm2 and so on.The common pressure differences are 1kPa, 1.23kpa, 1.47kpa, 10كيلو باسكال, etc.Theoretically, the flow rate is proportional to both area and pressure difference. For a numerical permeability sample, specifying the test area and pressure difference ensures that the flow rate is in a reasonable display range.For paper, due to the uneven and its own characteristics, في الحقيقة, the flow with area and differential pressure are not fully proportional to the relationship.So each test method standard specifies the test area and differential pressure.In special cases, such as the sample with too much or too little permeability, the test area can be appropriately reduced or increased to ensure the flow in a reasonable range of display.

5, several methods of paper permeability standards

في الوقت الحالي, the international standard methods for measuring paper permeability are as follows:① Schopper method,② Bendtsen method,③ Shffield method, ④ Gurley method,⑤ Oken method.

Each method has different requirements, but its core is to improve the accuracy of pressure difference and read flow rate in different ways within the measurement range.Shaub methodis to use the water level height difference to indicate the pressure difference on both sides of the test sample, in a certain time by measuring the drainage volume to calculate the flow.In this method, the difference of water level and height is used to indicate the pressure difference on both sides of the sample.It is not only inaccurate, but also limited in measuring range. This method is out of date now.Under constant pressure difference, “Bentsen method” و “Scherfeld methodswitch different range flowmeters according to different permeability of the sample. The flowmeter error is about 5%, which is difficult to improve.For low permeability samples, it is simply not suitable for an accurate flowmeter.AlthoughGelai methodbetter solve the measurement repeatability problem, but in its principle structure, the cylinder moves downward according to its own weight to form a pressure difference, and the pressure generated is affected by the falling speed and sealing fluid.There is also an error between the actual flow value and the displayed calculated value.The defects are particularly prominent for samples with higher permeability.

6, model RH-TQ100 permeability meter performance characteristics

Rh-tq100 paper permeability meter is developed and produced by Guangzhou Runhu Instrument Co., LTD.According to the different range of permeability of the sample, it is divided into different models A, B and C.Measuring range from 0.01μm/Pa·s to 3330μm/Pa·s, basically covering the whole range of paper permeability.

Rh-tq100 paper permeability tester adopts the latest mechanical and electrical technology, which greatly improves the test accuracy and accuracy.Pressure difference, the use of high precision pressure sensor, resolution of 0.001 كيلو باسكال;Flow measurement calculation, the use of mechatronics technology, very accurate measurement of the gas flow through the sample, its accuracy within 1%.Much higher than the general standard of 5% error.The measurement calibration of pressure difference and flow rate is traceable.Both high permeability paper (such as filter paper, molding paper, مناديل المراحيض, إلخ.) and low permeability paper (such as greaseproof paper, insulating paper, الورق المطلي, إلخ.), have a good measurement effect, measurement repeatability within 1%.The instrument has obtained two national patents.

 