SYD-510F1 Multi-functional Low Temperature Machine

مقدمة

This instrument is designed and manufactured in accordance with the requirements of the People’s Republic of China standard GB/T 510 “Method for Determination of the freezing point of Petroleum Products”, جيجابايت/ت 3535 “Method for determination of the Pour point of Petroleum”, جيجابايت/ت 6986 “Method for determination of the cloud point of petroleum”. It is suitable for the determination of the freezing point, pour point, turbidity point and other indexes of petroleum products according to the methods specified in the above standards.

With our LC-2 cold filter filter, the cold filter point of petroleum products can also be determined according to the method specified in SH/T 0248 “Determination of Cold filter Point of Diesel oil and domestic heating Oil”, the industry standard of the People’s Republic of China.

SYD-510F1 Multi-functional Low Temperature Machine

الميزات التقنية الرئيسية

1, the instrument adopts intelligent temperature control system, real-time display of cold tank temperature, temperature setting and control parameter adjustment is convenient, دقة التحكم في درجة الحرارة العالية.

2, multi-functional comprehensive, one machine can do pour point, cloud point, freezing point, cold filter point and many other tests.

3, the use of metal bath, refrigeration speed, درجة حرارة موحدة, eliminate the insecurity of liquid bath.

4, with automatic protection function, any cold tank high temperature abnormal, the instrument will automatically cut off the heating pipe power supply, to avoid damage to the instrument.

5, floor-to-ceiling structure, novel design, small footprint, with four wheels at the bottom, من السهل التحرك.

SYD-510F1 Multi-functional Low Temperature Machine

Main technical parameters and indicators

1, إمدادات الطاقة العاملة: تكييف (220±10%) الخامس; 50هرتز;

2. استهلاك الطاقة: ≤1700W

3, cold tank temperature control:

(1) Cold tank I: can be set -17℃~0 ℃, accuracy ±0.5 ℃, two bath isothermal;

(2) Cold tank II: set temperature to -34℃;

(3) Cold tank Ⅲ : can be set -51℃~-34℃, accuracy ±0.5 ℃, two bath isothermal;

(4) Cold tank Ⅳ : Set the temperature to -70℃.

4, refrigeration method: تبريد الضاغط;

5, البيئة القابلة للتطبيق: درجة حرارة: 15~28 ℃; الرطوبة النسبية: ≥85%;

6, الحجم الكلي: 810mm × 500mm × 840mm (الطول × العرض × الارتفاع)

7. وزن الآلة: 100كلغ.