Non-dispersive IR Gas Tester NDIR Non-Dispersive Infrared Gas Analyzer NDIR Multi-component Gas Test Machine
- وصف
- سؤال
وصف
Non-dispersive IR Gas Tester NDIR Non-Dispersive Infrared Gas Analyzer NDIR Multi-component Gas Test Machine
إنتاج
The LR-G105 infrared gas analyzer is an intelligent industrial online analyzer developed using the NDIR non-dispersive infrared principle combined with new microcomputer technology. The instrument is suitable for use in various non-explosion-proof environments and can analyze and monitor the content of total hydrocarbons, شركة, CO2, CH4, and other gases from ppm levels to 99.9% متصل.
The instrument features upper and lower limit alarms, digital and analog outputs, and customizable measurement ranges based on customer requirements and working conditions, providing cost-effective solutions.
Typical application areas
Petrochemical, steel and metallurgy, power generation, يبني, air separation, المواد الكيميائية, الأدوية, bio-fermentation, مواد بناء, مواد جديدة, وغيرها من الصناعات; including but not limited to online analysis and monitoring of CO, CO2, and other gas content in combustion processes, catalytic cracking reactions, various reaction vessels, صهاريج التخزين, process pipelines, combustion furnaces, and exhaust gases.
ميزات الصك
> Dot matrix LCD display, selectable display units: %, جزء في المليون, ملجم/م3
> Automatic real-time recording of measurement data (≤3000 entries), recorded data displayed in list or curve format
> Arbitrarily settable control points, passive relay output, standard 4-20mA and RS485
> Rich interface options, connectable to DCS and for communication with computers
> Gas path and program design with sensor sealed protection, ensuring stable overall performance
High cost-effectiveness; the core sensor uses imported products to ensure accurate analysis results, while other components use domestic parts of comparable quality to imported products, significantly reducing the overall price compared to imported instruments.
Non-dispersive IR Gas Tester NDIR Non-Dispersive Infrared Gas Analyzer NDIR Multi-component Gas Test Machine
المواصفات الفنية
| Measurable Gases | شركة, CO₂, CH₄, CnHm (other gases require further discussion) |
| Analysis Principle | Non-dispersive infrared |
| نطاق القياس | أ) شركة: 0-2000جزء في المليون, 0-2%, 0-5%, 0-10%, 0-20%, 0-50%, 0-100%
ب) CO₂: 0-2000جزء في المليون, 0-2%, 0-5%, 0-10%, 0-20%, 0-50%, 0-100% ج) CH₄: 0-2000جزء في المليون, 0-5%, 0-20%, 0-100% |
| دقة | ≤±1% F.S |
| التكرار | ≤±0.5% F.S |
| استقرار | ≤±0.5% F.S/7d |
| وقت الاستجابة | T90 ≤ 15 ثواني (depending on the principle and sampling distance) |
| بيئة التشغيل | درجة حرارة: -25℃~+55℃; Humidity ≤90%RH (غير التكثيف) |
| Sample Gas Pressure | 0.05 MPa ≤ Inlet pressure ≤ 0.1 MPa (outlet must be at atmospheric pressure) |
| Sample Gas Flow Rate | 400±10 ml/min |
| مزود الطاقة | 220بطالة ±10%, 50هرتز±5% (24V power supply requires special mention) |
| Data Logging | 3000 entries |
| انتاج | | 4~ 20mA, passive relay, RS485; can communicate with a computer |
| طريقة التشغيل | Button operation |
| أبعاد | 270مم (دبليو) × 134mm (ح) × 280mm (د) |
| Mounting Hole Size | 230مم (دبليو) × 134mm (ح) |
| وزن | تقريبا. 4كلغ |