Popis

ISO 10156 Analyzátor hořlavosti plynu Přístroj pro testování potenciálu požáru plynu Tester limitu výbušnosti plynu

Rozsah aplikace

1.1 Applicable to the determination of the explosion limits of flammable gases in air under normal pressure;

1.2 Applicable to the classification of flammable gas hazard levels;

1.3 Applicable to teaching, by studying the explosion limits of flammable gases, abstract concepts are made concrete and vivid, allowing students to understand the scientific rigor of chemical problems during the inquiry process, and enhancing their interest in chemistry.

 

Meet with the Normy

Meets the requirements of GB/T12474-2008 and ISO10156:2017 “Determination of Explosion Limits of Flammable Gases in Air”.

 

Main Design Features

3.1 Gas Proportioning: The instrument automatically performs gas proportioning through interface settings;

3.2 Fully Automatic Program Temperature Control: Utilizes P, já, and D functions for automatic temperature control;

3.3 Automatic Ignition Function

3.4 Automatic Calculation of Explosive Equivalent Based on Various Parameters;

3.5 Safe and Controllable Explosion Process with Complete Data Recording and Simultaneous Acquisition of Multiple Data Sources;

3.6 No Sparks Generated During Ignition and Explosion, Equipped with an Automatic Pressure Relief Protection Device;

3.7 Equipped with a Constant Temperature Chamber to Automatically Maintain a Constant Temperature in the Test Area;

3.8 Applicable to Explosion Experiment Research in Various Fields.

 

ISO 10156 Analyzátor hořlavosti plynu Přístroj pro testování potenciálu požáru plynu Tester limitu výbušnosti plynu

Technické specifikace

Composition of test apparatus Composed of reaction tube (explosion tube), ignition unit, stirring assembly, vacuum pump, tlakoměr, electromagnetic valve and other components
Reaction tube (explosion tube) struktura Made of cylindrical quartz tube; tube length: 1400mm±50mm, vnitřní průměr: 60mm±5mm, wall thickness over 2.0mm; a pressure relief valve with bore diameter over 25mm is installed at the tube bottom
Ignition system overall configuration Equipped with high-voltage electronic spark ignition device with adjustable ignition duration, electrode spacing and two ignition modes
High voltage ignition parameter 15kV high voltage electronic spark ignition, operating current 30mA
Adjustable ignition duration Ignition duration range: 0.1~1.0s, freely settable via touch screen
Electrode spacing range Discharge electrode spacing adjustable from 1mm to 6.5mm
Ignition operation mode Manual button ignition or automatic ignition optional
Electrode installation position Discharge electrodes are more than 100mm away from the reaction tube bottom and located at the geometric center of tube cross section
Stirring device structural design Adopts diaphragm pump for sealed circulating mixing with anti-explosion protection and pressure detection function
Stirring control function Stirring time freely configurable within 0~99.99min
Pump protection accessory Equipped with anti-explosion assembly to protect the diaphragm pump
Pressure measurement index Rozsah měření: 0-100kPa, rozlišení: 0.1kPa
Vacuum system composition Form negative pressure environment by rotary vane vacuum pump, with specified ultimate vacuum and tightness standard
Vacuum generation component Rotary vane vacuum pump for vacuum extraction
Ultimate vacuum degree Reachable vacuum level: 300Pa
Equipment tightness standard Pump out air to form 5mm water column inside the apparatus; pressure drop shall be less than 2mm water column 5min after pump shutdown
Constant temperature oven overall specification Independent temperature-controlled cabinet with fixed temperature and double-layer observation doors
Oven internal dimension Height 1.5m, width and depth each 0.5m
Constant temperature setting Stable operating temperature maintained at 50°C
Struktura dvouvrstvých dveří Two-layer door design: one layer of tempered glass, one layer of organic glass
Mixed gas distribution principle Partial pressure vacuum proportional gas distribution, fully automatic gas proportion matching by instrument program
Gas distribution channel layout Two gas inlet circuits for fuel gas and air respectively; volumetric flowmeters automatically adjust fuel-air mixing ratio
Gas pipeline assembly Pipeline system consists of vacuum pump, sample gas inlet, air inlet and air drying bottle
Explosion observation method Explosion occurrence judged manually through transparent viewing window
Safety protection configuration Double-layer anti-explosion shield with automatic pressure relief interlock function
Anti-explosion cover interlock function Explosion-proof cover fitted with transparent observation window and auto pressure relief valve; the valve automatically opens before ignition
Protective housing structure Double-layer protective shield installed for safety isolation
Apparatus calibration standard Calibrated with ethylene of purity ≥99.9% in accordance with Clause 7 of GB/T12474-2008
Calibration gas explosion limits Lower explosion limit of ethylene: 2.7%; upper explosion limit: 36%
Integrated test control system Adopts capacitive color LCD touch screen + PLC + PID temperature closed-loop control
Test repeatability error Repeatability deviation ≤5%
Test reproducibility error Reproducibility deviation ≤10%