FID Technology NMHC Analyzer Non-Methane Total Hydrocarbon Analyzer Total Hydrocarbon Concentration Tester
- Paglalarawan
- Pagtatanong
Paglalarawan
FID Technology NMHC Analyzer Non-Methane Total Hydrocarbon Analyzer Total Hydrocarbon Concentration Tester
Pagpapakilala sa Produkto
The LR-G103 Non-Methane Total Hydrocarbon Analyzer utilizes hydrogen flame ionization detection (FID) teknolohiya, offering high sensitivity and excellent stability. It features multiple selectable measurement ranges, capable of measuring trace hydrocarbons in high-purity gases to LEL (Lower Explosive Limit) levels in various gases. It is suitable for real-time analysis of total hydrocarbon concentration and lower explosive limit (LEL) in various organic gases.
Typical Application Areas
> Total hydrocarbon measurement in industrial process gases
> Gas concentration monitoring at the inlet and outlet of exhaust gas treatment devices (RTO/RCO/incinerators, atbp.)
> Real-time monitoring of total hydrocarbons in emission purification and chemical cleaning plants
> Monitoring of total hydrocarbon content in liquid oxygen of air separation units
> Real-time monitoring of organic compound concentration at the outlet of catalytic, biological, and activated carbon treatment devices
> Real-time monitoring of total hydrocarbons in crude coal gas conversion to clean gas
> Real-time monitoring of total hydrocarbons (or other organic compounds) in workshop, plant area, plant boundary, and ambient air
> Monitoring and early warning of organic compound leakage during production processes
> Real-time monitoring of total hydrocarbons in emission purification and chemical cleaning plants
> Monitoring of total hydrocarbon content in liquid oxygen of air separation units
Monitoring and early warning of the lower explosive limit of solvent-saturated air
Mga Tampok ng Produkto
> Fast response time: T90 < 2S, total response time < 5S
> Built-in heating pump, suitable for low-pressure sampling conditions
> The entire system is heated to prevent condensation and adsorption of organic compounds in the sample gas
> High measurement accuracy, better than 1% F.S.
> Can simultaneously output THC concentration and LEL value
> Built-in EPC flow control for stability
Response factors for various hydrocarbons are close to 1, resulting in stable and high-precision measurements
FID Technology NMHC Analyzer Non-Methane Total Hydrocarbon Analyzer Total Hydrocarbon Concentration Tester
Teknikal na Pagtutukoy
| Measured components | Total hydrogen, non-methane total hydrocarbons, explosive limit (LEL) |
| Detection range | THC: 0-50~100000 ppm; LEL: 0-100% |
| Limitasyon ng pagtuklas | 0.1 ppm po |
| Analyzer heating | 120℃~180℃ |
| Sampling flow rate | 0.5 SLM-1.5 SLM |
| Paulit ulit na pag uulit | ≤1% |
| Oras ng pagtugon | T90 < 2S, total response time < 5S |
| Zero drift | ≤±2%F.S./24h |
| Span drift | ≤±2%F.S./24h |
| Linearity error | ≤±1%F.S |
| Communication Protocol | Modbus communication protocol |
| Operating Environment | 5℃ ~40℃, 0~95% RH |
| Sample Gas Pressure | 80 kPa-110 kPa |
| Power Supply | 220 VAC, 50Hz |
| Hydrogen | Kadalisayan >99.99%, supplied from hydrogen cylinder or hydrogen generator, presyon 0.4 MPa |
| Air | Clean air, hydrocarbon concentration ≤0.3 mg/m³, presyon: 0.4 MPa |
| Pagmamaneho ng Gas | Nakumpres na hangin, free of water, alikabok na, and oil, presyon: 0.4 MPa |
| Calibration Gas | Methane/propane single or mixed calibration gas, 0.2 MPa |
| Instrument Power | Startup 500W, Running 150W |
| Mga Dimensyon ng Instrumento | 500 × 482 (435) × 213 mm |
| Timbang ng Instrumento | 21 kg |