説明

EN149 マスク難燃性能試験器

Testing standard:

Medical surgical mask

The flame retardant tester of surgical mask produced by our company is used for the flame retardant tester of surgical mask.Medical surgical mask meets the test standard yy0469-2011 EN149.

 

EN149 マスク難燃性能試験器

設備の特徴

  1. Automatic ignition.
  2. Digital display of flame temperature.
  3. Automatic movement for detection.

 

EN149 マスク難燃性能試験器

Technical parameters of detector

  1. Flame temperature: 800±50℃.
  2. The movement speed of the head mold is 60±5mm/s.
  3. The flame height is 40±4mm.
  4. The distance between the combustion top and the bottom of the sample is 20±2mm

 

EN149 マスク難燃性能試験器

商品名Mask flame retardant tester
基本的な原則The combustion test was carried out by adjusting the propane unit
炎の温度800±50℃
Time accuracy0.1s
test mediumdimethylmethane
制御モードutomatic control
適用範囲Surgical masks