説明

LR-230 気密漏れ検知器

使用

LR-230 series air-tightness detectoradopts the principle of direct pressure and uses air as the medium to pick up the leakage of the product. It is a non-destructive testing instrument.The working principle is to inflate the test product to the specified air pressure, then close the valve, after a period of stability, measure the change of air pressure during the test period, calculate the leakage through conversion.With its high precision and complete function, this instrument is widely used in sealing testing of open products or parts of closed products.

LR-230 気密漏れ検知器

機能的特徴

  • 7 inch touch screen display, Chinese and English language switch.
  • Large leakage, small leakage 2 alarm functions.
  • Alarm set,self-setting,Leakage pressure cap,Test pressure lower limit
  • Usb disk history storage, query, 印刷する, delete, screenshot
  • データストレージ,the instrument can store test results 24 hours a day
  • Test time can be set for each stage (0.1s-999s)
  • External control, do not need to use PLC program to achieve cylinder control

応用分野

Widely used in: electronic consumer goods, can only wear, wire material, 自動車, motorcycle industry, battery industry, valve pipe fittings, connection pipe, solenoid valve, lamps and lanterns, fire equipment, military supplies, household appliances and other fields

 

LR-230 気密漏れ検知器

技術仕様

モデルLR-230LF

Negative pressure Low pressure

LR-230MF

Negative pressure Middle pressure

 

LR-230F

Negative pressure

LR-230L

Low pressure

LR-230M

Middle pressure

LR-230H

High pressure

解決1pa5pa1pa1pa5pa20pa
Positive pressure test1~130kpa5~700kpaいいえ1~130kpa5~700kpa0.2~2Mpa
Negative pressure test-85~-1kpa-85~-1kpa-85~-1kpaNONONO
I/O Interface 7 input signals / 7 output signals
Test volumeThe test product volume is less than 10 リットル, the smaller the volume, the faster the test speed.
Historical dataHistorical data records show
試験時間Prepare1prepare 2pressurizedbalancedetectionexhaustremove

範囲:0.1s~~999.0s

サイズD310*W310*H210(んん)
重さAbout 8KG
AC 220V±10%,50/60HZ,電気の消費量:50バージニア州
テスト圧力源大流量と安定した圧力を備えたドライクリーンエア源
周囲温度と湿度温度の使用:5~40℃,保存温度 -20℃~70℃,80%RH以下,

でも結露はしない






    がカートに追加されました: