XE305S Series Semimicro Analytical/Precision Balance
- Beschrijving
- Navraag
Beschrijving
XE305S Series Semimicro Analytical/Precision Balance
Functies
ongeëvenaard sensorontwerp en verwerkingstechnologie, to ensure the sensitivity of trace weighing results
precisie: unieke supersingle-sensortechnologie, de nauwkeurigheid van de traceerweegresultaten garanderen, de stabiliteit en consistentie
snel: adopteer de sensortechnologie met één hendel, gecombineerd met een uitstekend circuitontwerp en software-algoritme, en voldoe aan de eis van het laboratorium om snel te wegen
design according to the American FDA data integrity principle, implement the data accurate, clear, easy to read, synchronization, original and traceability (end)
user management: through access control for user level 3 management.
electronic signature: electronic signature can be used to sign the final report of the weighing process.
audit tracking and weighing log: audit tracking is a time-stamped tamper-proof electronic log files, allows reconstruction and record to create, modify, delete the related events.
data transmission security: electronic records should be protected, to avoid being tampered with.
data output in line with the requirements of GLP/GMP
operational requirements is concise, clear, satisfy the user’s convenience, visual display good experience, at the same time, considering the special requirements of special industry users, such as non-contact operation at any time, any temperature can obtain accurate weighing results.
IsoCAL (the internal calibration and adjustment function, fully ensure accurate weighing results, using automatic temperature and time control of calibration and adjustment function.
Automatic identification of standard communication interface, and a printer
standard RS232, easy connection and peripherals.
GB/T26497-2022 electronic balance
JB/T9511-1999 balance packaging technology conditions
OIML R76-2 (2007 e), the European Union: the automatic weighing apparatus international recommendations
XE305S Series Semimicro Analytical/Precision Balance
Parameters
Model | XE305S | XE505S | XE805S | XE1205S | XE2205S |
maximaal bereik | 31G | 51G | 81G | 120 | 220 |
readability(d) | 0.01mg | ||||
repeatability(≤) Typical for 5% laden | ±0.04mg | ±0.05mg | ±0.06mg | ||
Herhaalbaarheid (≤) Typical value when the load is 100% | ±0.06mg | ±0.1mg | |||
linear error(≤) | ±0.05mg | ±0.08mg | ±0.1mg | ||
gevoeligheidsafwijking (+10℃-30℃) | 8 ppm/K | ||||
settling time | ≤6 second | ≤6 second | ≤6 second | ≤8 second | ≤10 second |
operating temperature range (OTR) | 18℃~23℃ | ||||
Operating humidity range | 50%RH~75%RH | ||||
Temperature compensating fluctuation | ≤1°C/time | ||||
Sample plate size | 85mm×85mm | ||||
grens dimensie | 240mm×430mm×360mm | ||||
Height of weighing room | 248mm | ||||
Startup preheating time | 30-60 seconde | ||||
working power supply | 220V±10% Allowable deviation rating, frequency: 50HZ±2% allowable deviation rating | ||||
Baud rate | 300、600、1200、2400、4800、9600 | ||||
expertise rating | Ⅰ level |