Programmable Comprehensive Electrical Safety Tester Hipot Insulation Ground Leakage Current Power Testing Machine Electrical Safety Compliance Test Instrument
- Beschrijving
- Navraag
Beschrijving
Programmable Comprehensive Electrical Safety Tester Hipot Insulation Ground Leakage Current Power Testing Machine Electrical Safety Compliance Test Instrument
Producteigenschappen
- It integrates AC/DC withstand voltage, isolatie, grounding, lekkage, and power tests, featuring fast testing speed and high efficiency.
- It uses a 5.6-inch TFT (640*480) display to set parameters and test parameters, with clear and rich display content.
- The tester can store test results in the form of an EXCEL table on a USB drive.
- Software upgrades can be performed via the USB drive.
- It adopts DDS digital signal synthesis technology to generate precise, stal, pure, and low-distortion sine waves.
- It allows adjustment of the rise and fall times of high voltage, enabling it to meet the requirements of different test objects.
- Two specific arc detection methods can be selected: current method and level method.
- Test results can be saved synchronously and support detailed and complete statistical operations.
- It supports online editing of test conditions by customers, facilitating intelligent and refined management.
- It has dual-frequency comprehensive testing, with a frequency range of 50 Hz and 60 Hz.
- It has an intuitive operation interface, supporting direct input with digital keys, dial input, and making operation more convenient.
- It has complete operation help prompts, which can effectively improve user efficiency.
- It has a bilingual operation interface, suitable for different user needs.
- It supports barcode scanner function.
- It comes with a PLC interface, RS232-interface, and optional LAN interface, RS485 interface, and USB interface.
Programmable Comprehensive Electrical Safety Tester Hipot Insulation Ground Leakage Current Power Testing Machine Electrical Safety Compliance Test Instrument
Technische specificatie
| Model | LR-CS9940 |
| Testmodi | AC Withstand / DC Withstand / Insulation Resistance / Ground Bond / Leakage Current / Power Test (AC/DC/IR/GR/LC/PW) |
| AC Withstand Voltage Test (5kV/20mA) | |
| Output Voltage Range | 0.250kV ~ 5.000kV |
| Voltage Accuracy | ±(2% lezing + 5V) |
| Voltage Resolution | 1V |
| Max Output Power | 100VA (5.000kV/20mA) |
| Max Rated Current | 20mA |
| Lower Current Limit Range | 0~20mA (0 = lower limit judgment disabled) |
| Current Ranges | 2mA, 20mA |
| Output Waveform | Pure Sine Wave |
| Waveform Distortion | ≤5% (No-load or pure resistive load) |
| Crest Factor | 1.3~1.5 |
| Signal Output Mode | Linear Power Amplifier |
| Voltage Ramp-up Time | 0, 0.3s~999.9s (0 = ramp-up function off) |
| Testduur | 0, 0.3s~999.9s (0 = continuous testing) |
| Voltage Ramp-down Time | 0, 0.3s~999.9s (0 = ramp-down function off) |
| Interval Time | 0.0s~999.9s (0 = interval function off) |
| Arc Detection Level | 0~9 (0 = arc detection disabled) |
| DC Withstand Voltage Test (6kV/10mA) | |
| Output Voltage Range | 0.250kV ~ 6.000kV |
| Voltage Accuracy | ±(2% lezing + 5V) |
| Voltage Resolution | 1V |
| Max Output Power | 60W (6.000kV/10mA) |
| Max Rated Current | 10mA |
| Lower Current Limit Range | 0~10mA (0 = lower limit judgment disabled) |
| Current Ranges | 2mA, 10mA |
| Ripple Factor | ≤5% (6kV/10mA) |
| Discharge Time | ≤200ms |
| Voltage Ramp-up Time | 0, 0.3s~999.9s (0 = ramp-up function off) |
| Testduur | 0, 0.3s~999.9s (0 = continuous testing) |
| Voltage Ramp-down Time | 0, 0.3s~999.9s (0 = ramp-down function off) |
| Interval Time | 0.0s~999.9s (0 = interval function off) |
| Insulation Resistance Test (DC 0.5kV / 1.0kV, Max 5GΩ) | |
| Test Voltage Options | 0.500kV / 1.000kV, Oplossing: 1V/Step |
| Upper Resistance Limit | 0MΩ~5000MΩ (0 = Disabled) |
| Lower Resistance Limit | 1MΩ~4999MΩ |
| Testduur | 0, 0.3s~999.9s (0 = continuous testing) |
| Interval Time | 0.0s~999.9s (0 = interval function off) |
| Voltage Ramp-up Time | 0, 0.3s~999.9s (0 = ramp-up function off) |
| Resistance Meter Resolution & Range Accuracy | 1MΩ~9.999MΩ: 0.001MΩ
10MΩ~99.99MΩ: 0.01MΩ 100MΩ~999.9MΩ: 0.1MΩ 1000MΩ~5000MΩ: 1MΩ 1MΩ~1000MΩ: ±5% 1000MΩ~5000MΩ: ±10% |
| Ground Bond Resistance Test (32A AC, Max 150mΩ) | |
| Output Current Range | 3.0A ~ 32.0A |
| Current Accuracy | ±(2% lezing + 0.1A) |
| Current Resolution | 0.1A |
| Max Output Power | 153.6VA |
| Upper Resistance Limit | Max. 510mΩ |
| Lower Resistance Limit | 0 ~ Upper Resistance Limit |
| Output Waveform | Pure Sine Wave |
| Waveform Distortion | ≤3% (No-load or pure resistive load) |
| Crest Factor | 1.3~1.5 |
| Output Drive Mode | Linear Power Amplifier |
| Current Ramp-up Time | 0, 0.3s~999.9s (0 = ramp-up function off) |
| Testduur | 0, 0.3s~999.9s (0 = continuous testing) |
| Interval Time | 0.0s~999.9s (0 = interval function off) |
| Leakage Current Test (External Power Supply, Max 3000VA Input) | |
| Output Voltage Range | 30.0V ~ 300.0V |
| Voltage Accuracy | ±(2% lezing + 1V) |
| Voltage Resolution | 0.1V |
| Current Ranges | 2mA, 20mA |
| Upper Current Limit | 0.001mA ~ 20.00mA |
| Lower Current Limit | 0 ~ Upper Current Limit |
| Human Body MD Network | Standard configuration complying with GB/T 12113 |
| Voltage Ramp-up Time | 0, 0.3s~999.9s (0 = ramp-up function off) |
| Testduur | 0, 0.3s~999.9s (0 = continuous testing) |
| Interval Time | 0.0s~999.9s (0 = interval function off) |
| Power Load Test (External Power Supply, Max 3000VA Input) | |
| Output Voltage Range | 30.0V ~ 300.0V |
| Voltage Accuracy | ±(2% lezing + 1V) |
| Voltage Resolution | 0.1V |
| Upper Current Limit | 0.100A ~ 14.00A |
| Lower Current Limit | 0 ~ 14.00A |
| Upper Power Limit | 1VA ~ 3000VA |
| Lower Power Limit | 0VA ~ 2999VA |
| Power Factor Range | 0.2 ~ 1.000 |
| Testduur | 0, 0.3s~999.9s (0 = continuous testing) |
| Interval Time | 0.0s~999.9s (0 = interval function off) |
| Meters Parameter | |
| Withstand Voltage Meter Range | 0.250kV ~ 6.000kV, RMS display
Nauwkeurigheid: ±(2% lezing + 5V), Oplossing: 1V |
| Withstand Current Meter | Bereik: 0.001mA ~ 20.00mA
Oplossing: 0.001mA (2mA range), 0.01mA (10mA/20mA range) Nauwkeurigheid: ±(2% lezing + 5 cijfers) |
| Ground Bond Current Meter | Bereik: 3.0A ~ 32.0A, RMS display
Oplossing: 0.1A, Nauwkeurigheid: ±(2% lezing + 2 cijfers) |
| Ground Resistance Meter | Bereik: 10.0mΩ ~ 510.0mΩ, 4-wire measurement
Oplossing: 1mΩ, Nauwkeurigheid: ±(2% lezing + 2mΩ) |
| Leakage Current Meter | Bereik: 0.1μA ~ 20.00mA
Oplossing: 0.001mA (2mA range), 0.01mA (20mA range) Nauwkeurigheid: ±(2% lezing + 5 cijfers) |
| Lekkage & Power Voltage Meter | Bereik: 30.0V~300.0V, RMS display
Oplossing: 0.1V, Nauwkeurigheid: ±(2% lezing + 1V) |
| Timer | Bereik: 0~999.9s, Oplossing: 0.1S, Nauwkeurigheid: ±(1% + 50mevrouw) |
| Opslag & Program Capacity | |
| Memory Groups | 50 sets |
| Test Steps per Group | 99 stappen |
| General Machine Parameters | |
| Input Power Supply | 220V±10%, 50Hz/60Hz±5% |
| Netto gewicht (without package) | Ongeveer. 22KG |
| Algemene dimensie (W×H×D) | 385mm × 135mm × 455mm |
| Werkomgeving | 0℃ ~ 40℃, 5%~75%RH, Low dust environment |