Opis

ASTM1238 ISO 1133 Plastic Melt Flow Index Testing Machine design and manufacture in accordance with GB/T 3682-2000 thermoplastic melt mass flow rate and melt volume flow rate Determination.

ISO 1133-2005 Plastics Determination of the melt mass-flow rate(MFR) and melt volume-flow rate(MVR) standardy.

It is not only suitable for polyethylene, polipropylen, polistyren, Żywica ABS o niższej temperaturze topnienia, takie jak test plastyczny, can also be applied to high temperature of polycarbonate, plastik fluorowy, nylon engineering plastics such as testing, therefore is widely used in plastic production, plastikowe produkty, petrochemical and other industries as well as the relevant universities and colleges, jednostki naukowo-badawcze.

Melt Flow Index Testing Machine Instrument features:

  1. LCD Englishdisplay and SMT keyboard input can directly and conveniently set time, temperature and sampling times.
  2. 32-bit single-chip microcomputer is adopted as the control kernel, with fast operation speed and high data accuracy.

3, using a 24-bit adc, temperature acquisition speed, acquisition accuracy, dokładna kontrola temperatury.

  1. Embedded operating system, reasonable internal task scheduling, accurate timing, delay and other control, to ensure the accuracy of temperature control, voice prompt and other operations.
  2. Can calculate, save and print test results.
  3. Adopt quality method (MFR) test.
  4. Automatic cutting during test and manual cutting during instrument adjustment.
  5. The machine has reasonable design, stable performance and convenient operation.

Melt Flow Index Testing Machine Main functions:

1, mass method melt flow rate test function.

2, automatic cutting function.

3.Melting index calculation function of mass method.

4.Melting index test result saving function.

5.Functions of saving results, consulting, printing and deleting

parametry techniczne:

  1. Extrude the material part

Discharging mouth diameter: Phi 2.095 ±0.005 mm

Length of discharge port: 8.000± 0.025 mm

Charging tube inner diameter: Phi 9.550 ± 0.025 mm

Length of charging barrel: 152± 0.1 mm

The piston rod head diameter: Φ 9.475± 0.015 mm

Długość główki tłoczyska: 6.350 0.100 mm

Effective stroke of piston rod: 30 mm

  1. Obciążenie testowe

The test load is obtained by combining different amounts of weights, which can constitute the nominal load required by the test. There are 8 grades in total:

Grade 1:0.325kg=(piston rod + weight tray + heat shield +1# weight body)= 3.187n

Grade 2:1.200kg(0.325kg+2#0.875kg weight)= 11.77n

Grade 3:2.160kg=(0.325kg+3#1.835kg weight)= 21.18n

Grade 4:3.800kg=(0.325kg+4#3.475kg weight)=37.26N

Grade 5:5.000kg=(0.325kg+5#4.675kg weight)=49.03N

Grade 6:10.000kg=(0.325kg+5#4.675k weight +6#5.000kg weight)=98.07N

Grade 7:12.500kg=(0.325kg+5#4.675kg weight +6#5.000kg weight +7#2.500kg weight)=122.58N

Grade 8:21.600 kg=(0.325kg+2#0.875kg weight +3#1.835kg weight +4#3.475kg weight +5#4.675kg weight +6#5.000kg weight +7#2.500kg weight +8#2.915kg weight)= 211.82n

  1. The range of temperature: temperatura pokojowa do 400 ℃
  2. The temperature control precision: plus lub minus 0.2 ℃
  3. Temperature fluctuations: plus lub minus 0.2 ℃
  4. The temperature resolution: 0.1 ℃
  5. Temperature recovery time: 4min
  6. Metoda badania: quality method (MFR)
  7. Zakres pomiaru: o. 1 ~ 300g / 10min(MFR)
  8. Dokładność pomiaru czasu: 0.1S (MFR, MVR)
  9. Metoda cięcia: automatic (podczas testu), podręcznik (when the instrument is adjusted)
  10. Cutting times: 9 czasy
  11. Cutting interval: 9999.9 sekundy
  12. Number of saved results: 20 copies
  13. Printer configuration: panel thermal printer
  14. Zasilacz: 220V, 50Hz
  15. Moc: 490W
  16. Overall dimensions: 290 X 410 X 630 (mm)

 