 Mask Synthetic Blood Penetration Test Machine for Blood Osmosis Test

Use

Mask synthetic blood penetration tester is used to determine the penetration performance of synthetic blood and other liquid-derived liquid jet, so as to evaluate its safety and protection performance.It is suitable for the determination of penetration of synthetic blood spillage in medical masks.It is widely used in the medical inspection department, safety inspection department and scientific research institute

 

Medical Mask Synthetic Blood Penetration Test Machine for Blood Osmosis Test

Executive standard:

YY 0469-2011: surgical masks for medical use

Gb19083-2010: technical requirements for medical protective masks;

YY/ t0691-2008: test method for the penetration of synthetic blood (fixed volume, horizontal injection) in the medical mask of protective equipment for infectious pathogens;

 

 Mask Synthetic Blood Penetration Test Machine for Blood Osmosis Test

Technical parameters

Test environment: temperature (21±5)℃;Relative humidity (85±10)%

Spray distance: (300±10)mm

Nozzle diameter: Φ 0.84 mm, 12.7 mm long

Liquid injection speed: 450cm/s;550 cm/s;635 cm/s

Spray pressure: standard 10.7kpa 16kpa 21.3kpa/adjustable

Power supply: AC (110~240) V 50Hz

Overall dimensions: 650×200×340 mm

 

Mask Synthetic Blood Penetration Test Machine for Blood Osmosis Test

Product features:

  1. The air pressure range can be set freely, and the setting range (0.5– 30)kPa;

2 touch screen Chinese display, friendly man-machine interface operation, aluminum alloy panel, equipped with metal keys, to ensure the durability of the instrument

  1. The sample table is made of 304 stainless steel, which is light, clean and never rusted.
  2. Constant pressure injection device, without manual adjustment

Gb19083-2010 technical requirements for medical protective masks

Synthetic blood penetration test;

2 mL synthetic blood was sprayed into the mask at a pressure of 10.7 kPa (80 mmHg). No infiltration should occur inside the mask.

YY 0469-2011 synthetic blood penetration test of surgical masks:

2 mL synthetic blood was sprayed into the mask at a pressure of 16.0 kPa (120 mmHg). No infiltration should occur inside the mask.

YY 0469-2011 synthetic blood penetration test of surgical masks:

Sample quantity: test with 3 samples.

Sample pretreatment: the sample was pretreated at least 4 h at temperature (21)C and relative humidity (85)%, and the test was carried out within 1 min after being taken out.

Test process: the sample is fixed on the sample fixture on the instrument, and the 2 mL table is placed 30.5cm away from the sample center

The surface tension was 0.002n /m with a pressure of 16.0 kPa (120 mmHg)

The needle tube with a diameter of 0.84mm was sprayed in a horizontal direction to the target area of the sample under test. Visual inspection was conducted within 10 s after removal.

Result treatment: check the inner side of the sample for penetration.If visual inspection is suspicious, use a absorbent cotton swab or similar to rub the inside of the target area and determine if there is synthetic blood penetration.The results shall meet the standard requirements.

 

 

 

 

 