Método de teste padrão ASTM F2013-2016 para determinação de acetaldeído residual em polímeros de frascos de tereftalato de polietileno usando um dispositivo automático de amostragem de headspace estático e cromatógrafo gasoso capilar com detector de ionização de chama

ASTM F2013-10 (2016) Standard Test Method for Determination of acetaldehyde Residues in polyethylene terephthalate Bottle Polymers Using an Automatic Static Headspace Sampling Device and Capillary Gas Chromatograph with a Flame Ionization Detector is especially used as a quality control tool for molding or synthesis operations.

Significado e uso

5.1 This test method is especially used as a quality control tool for molding or synthesis operations. Acetaldehyde is a volatile degradation product produced during PET melting processing. Como resultado, it is captured in the side wall of the molded product and slowly desorbed into the contents packaged therein. In some foods and beverages, AA can give off an undesirable taste, so it is important to understand its concentration in PET products used in food contact applications.

Método de teste padrão ASTM F2013-2016 para determinação de acetaldeído residual em polímeros de frascos de tereftalato de polietileno usando um dispositivo automático de amostragem de headspace estático e cromatógrafo gasoso capilar com detector de ionização de chama

5.2 Desorption conditions at 150°C for 60 minutes should ensure that no measurable AA is produced in the sample during desorption.

Etapa 1: Escopo

1.1 This test method covers gas chromatography for the determination of ppm residual acetaldehyde (AA) in polyethylene terephthalate (BICHO DE ESTIMAÇÃO) homopolymers and copolymers used in beverage bottle production. This includes sample types for amorphous and solid particle samples and billet samples, as opposed to bottling tests (test method D4509), which require 24 hours of desorption into the top space of the bottle at 23°C before concentration. Headspace was quantified by a similar GC method.

1.2 Os valores expressos em unidades SI serão considerados valores padrão. This standard does not include other units of measurement.

1.3 Esta norma não se destina a abordar todas as questões de segurança associadas ao seu uso. É responsabilidade do usuário desta norma estabelecer medidas apropriadas de segurança e saúde antes do uso e determinar a aplicabilidade das restrições regulatórias.

Método de teste padrão ASTM F2013-2016 para determinação de acetaldeído residual em polímeros de frascos de tereftalato de polietileno usando um dispositivo automático de amostragem de headspace estático e cromatógrafo gasoso capilar com detector de ionização de chama

2. Consulte o arquivo

Padrão ASTM

D4509 Test method for the determination of acetaldehyde content in the 24-hour gas (AIR) space in newly blown PET bottles

E691 Prática de realização de estudos interlaboratoriais para determinar a precisão dos métodos de teste

Compartilhe esta postagem