Descrição

Analisador de éter dimetílico de gás líquido GC-9860

Descrição

GC-9860 gas analyzer (all-in-one machine) is one of the typical domestic testing instruments, it uses the new technology of low power electronic information technology, the temperature control unit uses the intelligent digital display controller, equipped with a special thermal conductivity cell (TCD) detector and special liquid gas dimethyl ether detection filling column, the whole machine is highly integrated, controle de microcomputador, automatic parameter setting, Without manual setting, boot can work, to achieve high reliability and stability, operação simples, manutenção conveniente, widely used in liquefied gas production companies, liquefied gas storage and transportation stations, liquefied gas sales companies

Analisador de éter dimetílico de gás líquido GC-9860

Âmbito de aplicação

① Detection of added gases such as dimethyl ether, methanol, carbon monoxide and nitrogen in petroleum liquefied gas

② Detection of liquefied petroleum gas, natural gas, coal gas and other gas components

③ Determination of dimethyl ether purity

④ Detection of carbon dioxide, argon, methane, etc..

Analisador de éter dimetílico de gás líquido GC-9860

Main technical index

1. Main box temperature: temperatura do quarto +5 — 400℃ accuracy ±01℃

2. Thermal conductivity detector (TCD)

① Sensitivity: s≥3000mv.ml/mg(benzeno)

② Basic drift: 3%/h of full scale

Analisador de éter dimetílico de gás líquido GC-9860

Working condition

1, environmental temperature: 5℃-35℃

2, humidade relativa: less than 85%

3, power supply voltage: 220V±22V

4. Power supply frequency: 50HZ±0.5HZ

5, the maximum power consumption: 500C

6. There is no strong electromagnetic field interference, no corrosive gas, no strong vibration, no dramatic changes in indoor temperature, and no large convection in the air.

Analisador de éter dimetílico de gás líquido GC-9860

Technical performance

1, controle de temperatura: (automatic type: has been set, não há necessidade de mudar)

At room temperature add 5℃-400℃.

Precisão do controle de temperatura: no more than ±0.1℃.

2, detector parameters (automatic type: has been set, não há necessidade de mudar)

3, instrument size and weight:

Host size: 380×350×360 (comprimento × largura × altura) milímetros.

Peso: about 18kg.






    Foi adicionado ao seu carrinho:
    Confira