Descrição

Máquina de partículas de óleo LR-L100

Instrument overview

LR-L100 oil particle counter is an upgrade of L100 based on GB/T 18854-2002(ISO11171-1999) e outros padrões. Developed by the counting principle of light blocking method, it fully conforms to the corresponding national military standards, national standards and international standards. It can provide fast, preciso, reliable and repeatable test results and complete pollution monitoring and analysis report, which is suitable for the detection of particle pollution degree of hydraulic oil, lubricating oil, anti-fuel oil, insulating oil and turbine oil. It can be widely used in aviation, aeroespacial, electric power, petróleo, indústria química, transporte, porta, metalurgia, maquinaria, automobile manufacturing and other fields.

 

Máquina de partículas de óleo LR-L100

Característica principal

1.The counting principle of the light resistance (sombreamento) method formulated by the International hydraulic Standards Committee is adopted

2.Sensor laser de alta precisão, ampla faixa de teste, desempenho estável, barulho baixo, alta resolução

3.High pressure injection pump sampling method, can set the sample volume, stable injection speed, high sampling accuracy

4.The combination of positive and negative pressure injection system can achieve sample degassing, suitable for different viscosity test

5.Built-in pressure sensor, can set the pressure value, and automatically judge the air pressure in the cabin to ensure safety

6.Built-in air purification system to ensure that the test is not contaminated

7.Curva de calibração múltipla integrada, compatível com todos os padrões comuns no país e no exterior para calibração

8.GJB-420A integrado, GJB-420B, NAS1638, GB/T14039, ISO4406, SAE4059cpc, SAE4059F, SAE749D, γ T17216, QC/T29104, JB/T9737, DLT432, HH005-201 More than 10 commonly used standards, como 8, a test can give data results under all built-in standards

9.Supports custom standard tests and level determination

10.A maximum of 64 detection channels can be configured at the same time based on customer requirements

11.sistema de análise de dados integrado, one test can give all the standard test data and pollution levels

12.can carry out cleaning detection, pre-set the target cleaning requirements of particle concentration, real-time display and automatically stop when the requirements are met

13.Built-in viscosity, moisture and temperature sensor modules provide viscosity, moisture content saturation and ppm and temperature reference values while accurately measuring particle distribution (opcional)

14.can be arbitrarily set particle size, built-in nearly 10,000 particle size, easy to perform particle size analysis

15.A variety of sampling containers such as standard sampling bottles or cups can be used to meet the testing requirements of different industries

16.Full function automatic color touch screen operation, support Chinese and English dual interface, fácil de operar

17.can be single-channel and multi-channel calibration, to achieve automatic calibration function

18.possui interface RS232, pode ser conectado a um computador ou plataforma de laboratório para processamento de dados

19.provides mass data storage and printing functions, storing 1000 groups of data and supporting USB flash drives

 

Máquina de partículas de óleo LR-L100

Índice técnico

1.Fonte de luz: laser semicondutor

2.Faixa de tamanho de partícula: 0.8μm~600μm (de acordo com 1~100um ou 4~70um(c))

3.Detection channel: 8~64 channels Optional. Set any particle size

4.Sensibilidade: 0.8μm (ISO4402) or 3μm (c) (GB/T18854, ISO11171)

5.Resolução: < 10% (GB/T18854, ISO11171)

6.Repetibilidade: RSD<2%

7.Sample viscosity: ≤650cSt (excessive viscosity can be detected with heating or dilution method)

8.Volume de amostragem: 0.2~1000mL, intervalo: 0.1mL

9.Precisão de amostragem: better than ±0.5%

10.Velocidade de amostragem: 5 ~80mL/min

11.Pressure chamber: positive and negative pressure chamber device for sample degassing and high viscosity sample detection

12.Maximum vacuum of pressure chamber: -0.08MPa

13.Maximum positive pressure of the pressure chamber: 0.8MPa

14.Maximum particle concentration: 12000~40000 particles /mL

15.Temperatura (opcional) : Collection range: 1~100℃; Acquisition accuracy: 1℃

16.Water activity (opcional) : Collection range: 1~100%RH; Acquisition accuracy: 5%RH

17.Water content (opcional) : Collection range: 1~300ppm; Acquisition accuracy: 1ppm

18.Fonte de energia: 110~245V AC, 50Hz, 70C

19.Dimensão externa: 340mm╳410mm╳650mm

20.Peso líquido: 25kg






    Foi adicionado ao seu carrinho:
    Confira