Characteristics of network return control gas chromatograph:

  • Using the advanced 10/100m adaptive Ethernet communication interface and built-in IP protocol stack, the instrument can easily realize long-distance data transmission through Intranet and Internet of enterprises;It facilitates the erection of laboratory and simplifies the laboratory configuration and data management.
  • Low instrument internal design process of three separate connection, can connect to the local processing, unit director and supervisor, you can easily make unit director and supervisor real-time monitoring instrument is running and analysis data;
  • Low instrument equipped with built-in integration design return control chromatography workstation, get rid of the traditional data acquisition card processing, รองรับได้ถึง 253 sets of chromatographic work at the same time, data processing and the chromatograph, reached the domestic leading level;
  • The Modbus/TCP server built in the built-in integrated return control workstation can easily connect the analysis results to the DCS (distributed control system), and all parameters of the real-time return control chromatograph;
  • The instrument adopts vacuum fluorescence display technology, สูง – end atmosphere and convenient interface operation.Have two operating languages in English and Chinese to meet different user needs;
  • The instrument adopts the modular structure design, the design is clear, the replacement upgrade is convenient, protects the effectiveness of the investment;It can satisfy complex sample analysis, and it can be selected with many high performance detectors, such as FID, ทีซีดี, อีซีดี, FPD and NPD.
  • Can realize completely abandoned the traditional mechanical valve, pointer type pressure gauge mode!EPC technology is used to control gas path and improve the automation level of gas chromatography.
  • The real gas road failure automatic protection, pressure flow detection automatic ignition, automatic fire extinguishing automatic key, automatic gas path opening, reached a key start;
  • When the instrument is designed, it can easily complete the online analysis of the gas and liquid sample (with the sample parts).
  • The system is designed for automatic sampler interface, convenient for automatic sampler and future upgrade.

Main Specification

Operation display: vacuum fluorescent display (using agilent technology)

Temperature control area: 6

Low operating temperature: อุณหภูมิห้อง 4 ℃ ~ 450 ℃, the incremental: 1 ℃, precision: บวกหรือลบ 0.1 ℃

Program heating order number: 8-16 (ไม่จำเป็น)

Low Cheng Sheng rate: 0.1 ~ 50 ℃ / นาที (สามัญ);0.1 ~ 80 ℃ / นาที (ความเร็วสูง)

External events: 8 roads;Auxiliary control output 4

Type of sample: filling column, capillary injection sample, six-way valve gas injection, automatic head space sampling

Number of detectors: 3 (ขีดสุด);(เอฟไอดีเอ, ทีซีดี, อีซีดี, เอฟพีดี, NPD)

(1) it can be controlled by high-precision valve, and the pressure, real-time digital display of the flow and the PC synchronization display are realized.(2) an EPC control system with full back control;

Work mode of EPC and EFC: 2 types;Constant current mode, constant pressure mode

Work gas of EPC and EFC: 5 types;Nitrogen, hydrogen, อากาศ, helium, อาร์กอน

อีพีซี, EFC cheng sheng: 4

The control range of EPC and EFC:

ความดัน: 0 ~ 0.6mpa;Flow 0 ~ 100sccm or 0 ~ 500sccm (อากาศ)

Control accuracy of EPC and EFC: ความดัน 0.01 เคปาส;The flow rate of 0.01 SCCM

Low pressure sensor: ความแม่นยำ: full range of plus or minus 2% reproducibility: < + / – 0.05 KPa temperature coefficient: < + / – 0.01 เคปาส / ° C range: 0 ~ 0.3 MPa or 0 ~ 0.6 MPa;

Flow sensor: ความแม่นยำ: full range of < 5% reproducibility: < + 0.5% (full range) พิสัย: 0 ~ 500sccm

(คู่มือ, อัตโนมัติ, needle automatic detection) ไม่จำเป็น

อินเตอร์เฟซการสื่อสาร: Ethernet: IEEE802.3 hydrogen flame ionization detector (เอฟไอดีเอ)

ขีดจำกัดการตรวจจับ: Mt is less than 3 x 10-12g/s (positive hexadecaneisooctane solution);

เสียงรบกวนพื้นฐาน: น้อยกว่า 5 x 10-14a

การดริฟท์พื้นฐาน: ไม่เกิน 1 x 10-13a /30min

Linear range: มากกว่า 106

Highest use low temperature: 450 ℃ or less thermal conductivity detector (ทีซีดี)

The sensitivity: S is greater than 10000mV. Ml/mg (positive hexadecaneisooctane solution) (amplification 1, 2, 4, 8 times optional)

เสียงรบกวนพื้นฐาน: น้อยกว่า 10 mu V

การดริฟท์พื้นฐาน: ไม่เกิน 30 mu v/30min

Linear range: > 104 electronic capture detector (อีซีดี)

ขีดจำกัดการตรวจจับ: ไม่เกิน 1 x 10-14g/s (propane hexa6-isooctane solution)

เสียงรบกวนพื้นฐาน: less than 0.03mV

การดริฟท์พื้นฐาน: no more than 0.2mv /30min

Linear range: greater than or equal to 104

แหล่งที่มา: 63Ni flame photometric detector (เอฟพีดี)

ขีดจำกัดการตรวจจับ: (ส) น้อยกว่า 5 x 10-11 g/S;(ป) is less than or equal to 1 x 10-12g/s (เมทิลพาราไธออน – anhydrous ethanol solution)

เสียงรบกวนพื้นฐาน: less than 0.03mV

การดริฟท์พื้นฐาน: no more than 0.2mv /30min

Linear range: equal to 103 (ส) 102 (ป) nitrogen and phosphorus detector (NPD)

ขีดจำกัดการตรวจจับ: (เอ็น) is less than or equal to 1 times 10-13g/s;(ป) น้อยกว่า 5 x 10-14g/s (azobenz-malathionisooctane solution)

เสียงรบกวนพื้นฐาน: less than 0.03mV

การดริฟท์พื้นฐาน: no more than 0.2mv /30min

Linear range: greater than or equal to 102 (เอ็น) 103 (ป)

 

 