FID Technology NMHC Analyzer Non-Methane Total Hydrocarbon Analyzer Total Hydrocarbon Concentration Tester

Ознайомлення з продуктом

The LR-G103 Non-Methane Total Hydrocarbon Analyzer utilizes hydrogen flame ionization detection (FID) технології, offering high sensitivity and excellent stability. It features multiple selectable measurement ranges, capable of measuring trace hydrocarbons in high-purity gases to LEL (Lower Explosive Limit) levels in various gases. It is suitable for real-time analysis of total hydrocarbon concentration and lower explosive limit (LEL) in various organic gases.

Typical Application Areas

> Total hydrocarbon measurement in industrial process gases

> Gas concentration monitoring at the inlet and outlet of exhaust gas treatment devices (RTO/RCO/incinerators, тощо)

> Real-time monitoring of total hydrocarbons in emission purification and chemical cleaning plants

> Monitoring of total hydrocarbon content in liquid oxygen of air separation units

> Real-time monitoring of organic compound concentration at the outlet of catalytic, біологічний, and activated carbon treatment devices

> Real-time monitoring of total hydrocarbons in crude coal gas conversion to clean gas

> Real-time monitoring of total hydrocarbons (or other organic compounds) in workshop, plant area, plant boundary, and ambient air

> Monitoring and early warning of organic compound leakage during production processes

> Real-time monitoring of total hydrocarbons in emission purification and chemical cleaning plants

> Monitoring of total hydrocarbon content in liquid oxygen of air separation units

Monitoring and early warning of the lower explosive limit of solvent-saturated air

Особливості продукту

> Fast response time: Т90 < 2С, total response time < 5С

> Built-in heating pump, suitable for low-pressure sampling conditions

> The entire system is heated to prevent condensation and adsorption of organic compounds in the sample gas

> Висока точність вимірювань, Краще, ніж 1% F.S.

> Can simultaneously output THC concentration and LEL value

> Built-in EPC flow control for stability

Response factors for various hydrocarbons are close to 1, resulting in stable and high-precision measurements

 

FID Technology NMHC Analyzer Non-Methane Total Hydrocarbon Analyzer Total Hydrocarbon Concentration Tester

Технічна специфікація

Measured components Total hydrogen, non-methane total hydrocarbons, explosive limit (LEL)
Дальність виявлення THC: 0-50~100000 ppm; LEL: 0-100%
Межа виявлення 0.1 ppm
Analyzer heating 120℃~180℃
Sampling flow rate 0.5 SLM-1.5 SLM
Повторюваність ≤1%
Час відповіді Т90 < 2С, total response time < 5С
Zero drift ≤±2%F.S./24h
Span drift ≤±2%F.S./24h
Linearity error ≤±1%F.S
Протокол зв'язку Modbus communication protocol
Операційне середовище 5℃ ~40℃, 0~95% RH
Sample Gas Pressure 80 kPa-110 kPa
Джерело живлення 220 VAC, 50Гц
водень Чистота >99.99%, supplied from hydrogen cylinder or hydrogen generator, тиск 0.4 МПа
Повітря Clean air, hydrocarbon concentration ≤0.3 mg/m³, тиск: 0.4 МПа
Driving Gas Стиснене повітря, free of water, пил, and oil, тиск: 0.4 МПа
Calibration Gas Methane/propane single or mixed calibration gas, 0.2 МПа
Instrument Power Startup 500W, Running 150W
Розміри приладу 500 × 482 (435) × 213 мм
Вага приладу 21 кг