설명

1.측정 범위: 0—1250um or 0—50mil

  1. 정확성:± [(1%~3%)H+1um] 메모:H는 두께 판독값입니다.
  2. powers down after 60 seconds of no activity.

    1、Power-on
    Keystroke onceWait until READY! is displayed on the LCD. Now the gauge is ready for measurements.
    2、Zeroing
    Keystroke twice; when BASE ZERO is displayed, place flat on surface; hold steady. After measurement is obtained, DONE is shown and gauge beeps. Then lift gauge.
    메모: Please conduct zeroing process after changing base. Zeroing process could be repeated for better results.
    3、Two-Point User Calibration
    Keystroke three times, enter into the process of tow-point user calibration; 언제
    USER CAL BASE is shown on the LCD, please conduct a measurement on the base first; 언제
    S1 is shown on the LCD, place Shim S1 on the base, conduct a measurement on shim S1 and lift gauge; 언제
    S2 is shown on the LCD, place Shim S2 on the base, conduct a measurement on shim S2 and lift gauge; When DONE is displayed, it is ready for use.
    메모: The two-point user calibration process could be repeated for better results.
    4、Unit Change
    Keystroke four times, convert from ums to mils or vice versa.
    5、Reset
    Keystroke five times, gauge resets.






    장바구니에 추가되었습니다:
    점검