Etapy kalibracji maszyny do badania zużycia TABER nie są trudne

TABER wear testing machine introduction:
This machine is used for wear resistance test of leather, płótno, farba, papier, płytka podłogowa, sklejka, szkło, natural rubber, itp. The method is to cut the specimen with a standard cutter, wear it with a specified type of grinding wheel, load it with a certain weight, and take out the specimen after the specified number of revolutions, observe the condition of the specimen or compare the weight difference before and after the test. Mainly test the wear resistance of the material, wear resistance with two standard small grinding wheels, calculate the importance of the material and see if there is damage after grinding.
Mainly used in leather, skóra, płótno, farba, papier, floor, sklejka, szkło, natural rubber suitcase, carpet, odzież, szkło, plastic coating, ceramic tile, metal coating, farba, lakier, decorative sheet, high pressure sheet, Plastikowy, włókienniczy, elastic floor mat, traffic paint, anodized layer, blankets, części elektroniczne, decorative board, wax, etykiety, leather cases, dental materials, Automotive interior decoration, żywica, furniture and research laboratories; Commodity inspection arbitration, działy nadzoru technicznego, itp. to detect the performance of leather products.

TABER wear testing machine calibration steps:
Krok 1, set the rated revolution of the counter to 60r/min:
Krok 2, press the RST key to clear the counter;
Step 3,Press the start button and start the stopwatch at the same time.
Krok 4, the counter records to 60 obr./min. The testing machine automatically stops, stops the stopwatch time;
Krok 5, Record stopwatch time:
Repeat steps 2 Do 5 powyżej, both times should be within the range of 60s±2s.