Description

Programmable Comprehensive Electrical Safety Tester Hipot Insulation Ground Leakage Current Power Testing Machine Electrical Safety Compliance Test Instrument

Product Features

  1.  It integrates AC/DC withstand voltage, insulation, grounding, leakage, and power tests, featuring fast testing speed and high efficiency.
  2.  It uses a 5.6-inch TFT (640*480) display to set parameters and test parameters, with clear and rich display content.
  3.  The tester can store test results in the form of an EXCEL table on a USB drive.
  4.  Software upgrades can be performed via the USB drive.
  5.  It adopts DDS digital signal synthesis technology to generate precise, stable, pure, and low-distortion sine waves.
  6.  It allows adjustment of the rise and fall times of high voltage, enabling it to meet the requirements of different test objects.
  7.  Two specific arc detection methods can be selected: current method and level method.
  8.  Test results can be saved synchronously and support detailed and complete statistical operations.
  9.  It supports online editing of test conditions by customers, facilitating intelligent and refined management.
  10.  It has dual-frequency comprehensive testing, with a frequency range of 50 Hz and 60 Hz.
  11.  It has an intuitive operation interface, supporting direct input with digital keys, dial input, and making operation more convenient.
  12.  It has complete operation help prompts, which can effectively improve user efficiency.
  13.  It has a bilingual operation interface, suitable for different user needs.
  14.  It supports barcode scanner function.
  15.  It comes with a PLC interface, RS232 interface, and optional LAN interface, RS485 interface, and USB interface.

 

Programmable Comprehensive Electrical Safety Tester Hipot Insulation Ground Leakage Current Power Testing Machine Electrical Safety Compliance Test Instrument

Technical Specification

Model LR-CS9940
Test Modes AC Withstand / DC Withstand / Insulation Resistance / Ground Bond / Leakage Current / Power Test (AC/DC/IR/GR/LC/PW)
AC Withstand Voltage Test (5kV/20mA)  
Output Voltage Range 0.250kV ~ 5.000kV
Voltage Accuracy ±(2% reading + 5V)
Voltage Resolution 1V
Max Output Power 100VA (5.000kV/20mA)
Max Rated Current 20mA
Lower Current Limit Range 0~20mA (0 = lower limit judgment disabled)
Current Ranges 2mA, 20mA
Output Waveform Pure Sine Wave
Waveform Distortion ≤5% (No-load or pure resistive load)
Crest Factor 1.3~1.5
Signal Output Mode Linear Power Amplifier
Voltage Ramp-up Time 0, 0.3s~999.9s (0 = ramp-up function off)
Test Duration 0, 0.3s~999.9s (0 = continuous testing)
Voltage Ramp-down Time 0, 0.3s~999.9s (0 = ramp-down function off)
Interval Time 0.0s~999.9s (0 = interval function off)
Arc Detection Level 0~9 (0 = arc detection disabled)
DC Withstand Voltage Test (6kV/10mA)  
Output Voltage Range 0.250kV ~ 6.000kV
Voltage Accuracy ±(2% reading + 5V)
Voltage Resolution 1V
Max Output Power 60W (6.000kV/10mA)
Max Rated Current 10mA
Lower Current Limit Range 0~10mA (0 = lower limit judgment disabled)
Current Ranges 2mA, 10mA
Ripple Factor ≤5% (6kV/10mA)
Discharge Time ≤200ms
Voltage Ramp-up Time 0, 0.3s~999.9s (0 = ramp-up function off)
Test Duration 0, 0.3s~999.9s (0 = continuous testing)
Voltage Ramp-down Time 0, 0.3s~999.9s (0 = ramp-down function off)
Interval Time 0.0s~999.9s (0 = interval function off)
Insulation Resistance Test (DC 0.5kV / 1.0kV, Max 5GΩ)  
Test Voltage Options 0.500kV / 1.000kV, Resolution: 1V/Step
Upper Resistance Limit 0MΩ~5000MΩ (0 = Disabled)
Lower Resistance Limit 1MΩ~4999MΩ
Test Duration 0, 0.3s~999.9s (0 = continuous testing)
Interval Time 0.0s~999.9s (0 = interval function off)
Voltage Ramp-up Time 0, 0.3s~999.9s (0 = ramp-up function off)
Resistance Meter Resolution & Range Accuracy 1MΩ~9.999MΩ: 0.001MΩ

10MΩ~99.99MΩ: 0.01MΩ

100MΩ~999.9MΩ: 0.1MΩ

1000MΩ~5000MΩ: 1MΩ

1MΩ~1000MΩ: ±5%

1000MΩ~5000MΩ: ±10%

Ground Bond Resistance Test (32A AC, Max 150mΩ)  
Output Current Range 3.0A ~ 32.0A
Current Accuracy ±(2% reading + 0.1A)
Current Resolution 0.1A
Max Output Power 153.6VA
Upper Resistance Limit Max. 510mΩ
Lower Resistance Limit 0 ~ Upper Resistance Limit
Output Waveform Pure Sine Wave
Waveform Distortion ≤3% (No-load or pure resistive load)
Crest Factor 1.3~1.5
Output Drive Mode Linear Power Amplifier
Current Ramp-up Time 0, 0.3s~999.9s (0 = ramp-up function off)
Test Duration 0, 0.3s~999.9s (0 = continuous testing)
Interval Time 0.0s~999.9s (0 = interval function off)
Leakage Current Test (External Power Supply, Max 3000VA Input)  
Output Voltage Range 30.0V ~ 300.0V
Voltage Accuracy ±(2% reading + 1V)
Voltage Resolution 0.1V
Current Ranges 2mA, 20mA
Upper Current Limit 0.001mA ~ 20.00mA
Lower Current Limit 0 ~ Upper Current Limit
Human Body MD Network Standard configuration complying with GB/T 12113
Voltage Ramp-up Time 0, 0.3s~999.9s (0 = ramp-up function off)
Test Duration 0, 0.3s~999.9s (0 = continuous testing)
Interval Time 0.0s~999.9s (0 = interval function off)
Power Load Test (External Power Supply, Max 3000VA Input)  
Output Voltage Range 30.0V ~ 300.0V
Voltage Accuracy ±(2% reading + 1V)
Voltage Resolution 0.1V
Upper Current Limit 0.100A ~ 14.00A
Lower Current Limit 0 ~ 14.00A
Upper Power Limit 1VA ~ 3000VA
Lower Power Limit 0VA ~ 2999VA
Power Factor Range 0.2 ~ 1.000
Test Duration 0, 0.3s~999.9s (0 = continuous testing)
Interval Time 0.0s~999.9s (0 = interval function off)
Meters Parameter  
Withstand Voltage Meter Range 0.250kV ~ 6.000kV, RMS display

Accuracy: ±(2% reading + 5V), Resolution: 1V

Withstand Current Meter Range: 0.001mA ~ 20.00mA

Resolution: 0.001mA (2mA range), 0.01mA (10mA/20mA range)

Accuracy: ±(2% reading + 5 digits)

Ground Bond Current Meter Range: 3.0A ~ 32.0A, RMS display

Resolution: 0.1A, Accuracy: ±(2% reading + 2 digits)

Ground Resistance Meter Range: 10.0mΩ ~ 510.0mΩ, 4-wire measurement

Resolution: 1mΩ, Accuracy: ±(2% reading + 2mΩ)

Leakage Current Meter Range: 0.1μA ~ 20.00mA

Resolution: 0.001mA (2mA range), 0.01mA (20mA range)

Accuracy: ±(2% reading + 5 digits)

Leakage & Power Voltage Meter Range: 30.0V~300.0V, RMS display

Resolution: 0.1V, Accuracy: ±(2% reading + 1V)

Timer Range: 0~999.9s, Resolution: 0.1s, Accuracy: ±(1% + 50ms)
Storage & Program Capacity  
Memory Groups 50 sets
Test Steps per Group 99 steps
General Machine Parameters  
Input Power Supply 220V±10%, 50Hz/60Hz±5%
Net Weight (without package) Approx. 22KG
Overall Dimension (W×H×D) 385mm × 135mm × 455mm
Operating Environment 0℃ ~ 40℃, 5%~75%RH, Low dust environment