- 描述
- 询问
描述
1.测量范围: 0—1250um 或 0—50mil
- 准确性:± [(1%〜3%)H+1um] 笔记:H是厚度读数
- 之后断电 60 无活动秒数.
1、Power-on
按键一次;等待 LCD 上显示 READY!. 现在仪表已准备好进行测量.
2、归零
按键两次; 当显示 BASE ZERO 时, place flat on surface; 持稳. 获得测量值后, 显示“DONE”(完成)并发出蜂鸣声. 然后举升表.
笔记: 更换底座后请进行调零处理. 可以重复调零过程以获得更好的结果.
3、Two-Point User Calibration
按键三下, enter into the process of tow-point user calibration; when
USER CAL BASE is shown on the LCD, please conduct a measurement on the base first; when
S1 is shown on the LCD, place Shim S1 on the base, conduct a measurement on shim S1 and lift gauge; when
S2 is shown on the LCD, place Shim S2 on the base, conduct a measurement on shim S2 and lift gauge; When DONE is displayed, it is ready for use.
笔记: The two-point user calibration process could be repeated for better results.
4、单位变更
击键四次, 从 um 转换为 mil,反之亦然.
5、重置
Keystroke five times, gauge resets.