ASTM D 3278-2021 Pagpapasiya ng Liquid Flash Point na may isang Maliit na Closed Cup Apparatus

Kahulugan at Layunin
5.1 Flash point measurement Sample tugon sa init at pag aapoy pinagmulan sa ilalim ng kinokontrol na mga kondisyon ng laboratoryo. It is only one of many characteristics to be considered when assessing the overall flammability hazard of a material.

5.2 Flash points are used by government regulators in transportation and safety regulations to define and classify flammable and combustible materials. For a precise definition of these classes, see the specific regulations involved.

5.3 Flash points can indicate the possible presence of highly volatile and flammable impurities or contaminants in a given liquid, such as the presence of residual solvents in solvent refined dry oils.

5.4 These balanced flash point test methods use smaller specimens (2 mL) and shorter test times (1 min) compared to conventional non-equilibrium test methods such as test Method D56 and Test Method D93.

5.5 Test method D3828, Test method D8174 and ISO 3679 are similar test methods and use the same equipment.

ASTM D 3278-2021 Pagpapasiya ng Liquid Flash Point na may isang Maliit na Closed Cup Apparatus

Saklaw ng
1.1 These test methods include procedures to determine whether A material flashes at a specified temperature when using a small closed cup device (flash/no flash method A) or to determine the minimum finite temperature at which the material flashes (Paraan B). The test method is applicable to paints, enamels, lacquers, magbarnis, solvents and related products with flash points between 0°C and 110°C (32°F and 230°F) and viscosity below 15 000 mm 2/ pangalawa (cSt) at 25 °C (77 °F).

Tala 1: Tests can be performed at higher or lower temperatures, but the accuracy has not been determined.

Tala 2: More viscous materials can be tested according to Appendix A4.

Tala 3: Organic peroxides can be tested in accordance with Appendix A5, which describes the applicable safety precautions.

Tala 4: The United States Department of Labor (OSHA, Hazard Communications), the United States Department of Transportation (RSPA) and the United States Environmental Protection Agency (EPA) have designated test method D3278 as one of several acceptable methods for determining the flash point of liquids in their regulations.

ASTM D 3278-2021 Pagpapasiya ng Liquid Flash Point na may isang Maliit na Closed Cup Apparatus

1.2 Values expressed in SI units should be considered standard values. The values given in parentheses following SI units are for reference only and are not considered standard values.

1.3 Ang pamantayang ito ay ginagamit upang sukatin at ilarawan ang reaksyon ng mga materyales, mga produkto o mga bahagi sa init at apoy sa ilalim ng kinokontrol na mga kondisyon, ngunit hindi ba sa sarili nito ay naglalaman ng lahat ng mga kadahilanan na kinakailangan para sa panganib ng sunog o sunog panganib pagtatasa ng mga materyales, mga produkto o mga bahagi sa ilalim ng aktwal na mga kondisyon ng sunog.

1.4 Ang pamantayang ito ay hindi nilayon upang matugunan ang lahat ng mga isyu sa kaligtasan, kung meron man, na nauugnay sa paggamit nito. Responsibilidad ng gumagamit ng pamantayang ito na magtatag ng angkop na kaligtasan, mga kasanayan sa kalusugan at kapaligiran at upang matukoy ang pagiging angkop ng mga paghihigpit sa regulasyon bago gamitin.

1.5 Ang internasyonal na Pamantayang ito ay binuo alinsunod sa mga internasyonal na kinikilalang prinsipyo ng standardisasyon na itinatag sa Desisyon sa mga Prinsipyo para sa Pag unlad ng Internasyonal na Pamantayan, Mga Patnubay at Rekomendasyon na inilabas ng Komite sa mga Teknikal na Hadlang sa Kalakalan ng World Trade Organization.