Description

GC-9860 Liquid Gas Dimethyl Ether Analyzer

Description

GC-9860 gas analyzer (all-in-one machine) is one of the typical domestic testing instruments, it uses the new technology of low power electronic information technology, the temperature control unit uses the intelligent digital display controller, equipped with a special thermal conductivity cell (TCD) detector and special liquid gas dimethyl ether detection filling column, the whole machine is highly integrated, microcomputer control, automatic parameter setting, Without manual setting, boot can work, to achieve high reliability and stability, simple operation, convenient maintenance, widely used in liquefied gas production companies, liquefied gas storage and transportation stations, liquefied gas sales companies

GC-9860 Liquid Gas Dimethyl Ether Analyzer

Scope of application

① Detection of added gases such as dimethyl ether, methanol, carbon monoxide and nitrogen in petroleum liquefied gas

② Detection of liquefied petroleum gas, natural gas, coal gas and other gas components

③ Determination of dimethyl ether purity

④ Detection of carbon dioxide, argon, methane, etc.

GC-9860 Liquid Gas Dimethyl Ether Analyzer

Main technical index

1. Main box temperature: room temperature +5 — 400℃ accuracy ±01℃

2. Thermal conductivity detector (TCD)

① Sensitivity: s≥3000mv.ml/mg(benzene)

② Basic drift: 3%/h of full scale

GC-9860 Liquid Gas Dimethyl Ether Analyzer

Working condition

1, environmental temperature: 5℃-35℃

2, relative humidity: less than 85%

3, power supply voltage: 220V±22V

4. Power supply frequency: 50HZ±0.5HZ

5, the maximum power consumption: 500W

6. There is no strong electromagnetic field interference, no corrosive gas, no strong vibration, no dramatic changes in indoor temperature, and no large convection in the air.

GC-9860 Liquid Gas Dimethyl Ether Analyzer

Technical performance

1, temperature control: (automatic type: has been set, no need to change)

At room temperature add 5℃-400℃.

Temperature control accuracy: no more than ±0.1℃.

2, detector parameters (automatic type: has been set, no need to change)

3, instrument size and weight:

Host size: 380×350×360 (length × width × height) mm.

Weight: about 18kg.






    has been added to your cart: