Description

AHS-2036 Fully Automatic Headspace Sampler

Introduction of the instrument:

AHS-2036 Full automatic headspace sampler (36 bits) is a new product of our company with fully automated design, unattended, touch large screen display, which can realize multiple sample overlapping heating. Boot system self-check, fault alarm and prompt function, temperature overload protection function, leakage protection function and other beautiful appearance, more convenient operation of the new generation of automatic headspace sampler.

AHS-2036 Fully Automatic Headspace Sampler

Features:

▲ 7.0 inch touch LCD screen, simple operation, one button start; It does not need to spend too much energy to get started, convenient for users to operate quickly.

▲ Intelligent automation: automatic heating balance, pressure, sampling, sampling, analysis and blowing after analysis and other functions, automatic replacement of sample bottles to achieve full automation process.

▲ Preset 36 samples, can realize sample overlapping heating, greatly improve the utilization of time. The reliability and consistency of analysis work are improved without manual intervention.

▲ With complete parameter setting, real-time working state, running time, can realize the same sample multiple sampling analysis.

▲ Comprehensive power-on self-test function, sample plate automatic zero positioning, component failure, wiring failure and other automatic alarm; Temperature overload protection function.

▲ Multiple sampling modes: single bottle multiple times, single bottle single time, multiple bottles single time.

▲ Using the classic positive pressure sampling method, you can determine liquid or solid samples, high detection sensitivity. Normal pressure injection without baseline drift.

▲ Sample transmission pipeline using inert material, sampling valve and transmission pipeline connection point in the heating incubator, no transmission cold point, to ensure the integrity of the sample.

▲ All actions are driven by motor, without external driving gas, safe and convenient to use.

▲ The sampling system has automatic backblowing function, which can effectively avoid cross contamination of different samples (patent number: ZL 2021 2 1874159.8);

Real-time display of working status, running time, pressure, etc.

▲ Can synchronously start GC, chromatographic data processing workstation, can also realize the reverse control between GC;

▲ The instrument is equipped with a variety of imported instruments special interface and injection needle protection equipment (patent number: ZL 2021 2 1874184.6), can be connected to all models at home and abroad

AHS-2036 Fully Automatic Headspace Sampler

Technical parameters:

Sample heating range: standard 40℃ — 280℃, temperature control accuracy 1℃

● Sampling valve heating range: standard 40℃ – 220℃, temperature control accuracy 1℃

(According to customer requirements, can be configured to 300℃)

● Sample transfer tube heating range: standard 40℃ – 220℃, temperature control accuracy 1℃

● Temperature accuracy: 0.5%

● Temperature stability: <±0.5℃

● Pressure time: 0-9999s

● Sampling time: 0-30min

● Sampling time: 0-9999s

● Cleaning time: 0-30min

● Pressure: 0~0.25Mpa (continuously adjustable)

● Volume of quantitative tube: 1ml (other specifications can be customized, such as 0.5ml, 2ml, 5ml, etc.)

● Headspace bottle specifications: 10ml or 20ml (other specifications can be customized, such as 50ml, 100ml, etc.)

● Sample station: 36  RSD: <1.0% (100ppm ethanol solution)

● Backblow cleaning flow: 0 ~ 100ml/min (continuously adjustable)

● Synchronously start chromatographic data processing workstation, GC and other devices