LR-L100 Oil Particle Machine

Instrument overview

LR-L100 oil particle counter is an upgrade of L100 based on GB/T 18854-2002(ISO11171-1999) and other standards. Developed by the counting principle of light blocking method, it fully conforms to the corresponding national military standards, national standards and international standards. It can provide fast, accurate, reliable and repeatable test results and complete pollution monitoring and analysis report, which is suitable for the detection of particle pollution degree of hydraulic oil, lubricating oil, anti-fuel oil, insulating oil and turbine oil. It can be widely used in aviation, aerospace, electric power, petroleum, chemical industry, transportation, port, metallurgy, machinery, automobile manufacturing and other fields.

 

LR-L100 Oil Particle Machine

Main feature

1.The counting principle of the light resistance (shading) method formulated by the International hydraulic Standards Committee is adopted

2.High precision laser sensor, wide test range, stable performance, low noise, high resolution

3.High pressure injection pump sampling method, can set the sample volume, stable injection speed, high sampling accuracy

4.The combination of positive and negative pressure injection system can achieve sample degassing, suitable for different viscosity test

5.Built-in pressure sensor, can set the pressure value, and automatically judge the air pressure in the cabin to ensure safety

6.Built-in air purification system to ensure that the test is not contaminated

7.Built-in multiple calibration curve, compatible with all common standards at home and abroad for calibration

8.Built-in GJB-420A, GJB-420B, NAS1638, GB/T14039, ISO4406, SAE4059cpc, SAE4059F, SAE749D, γ T17216, QC/T29104, JB/T9737, DLT432, HH005-201 More than 10 commonly used standards, such as 8, a test can give data results under all built-in standards

9.Supports custom standard tests and level determination

10.A maximum of 64 detection channels can be configured at the same time based on customer requirements

11.built-in data analysis system, one test can give all the standard test data and pollution levels

12.can carry out cleaning detection, pre-set the target cleaning requirements of particle concentration, real-time display and automatically stop when the requirements are met

13.Built-in viscosity, moisture and temperature sensor modules provide viscosity, moisture content saturation and ppm and temperature reference values while accurately measuring particle distribution (optional)

14.can be arbitrarily set particle size, built-in nearly 10,000 particle size, easy to perform particle size analysis

15.A variety of sampling containers such as standard sampling bottles or cups can be used to meet the testing requirements of different industries

16.Full function automatic color touch screen operation, support Chinese and English dual interface, easy to operate

17.can be single-channel and multi-channel calibration, to achieve automatic calibration function

18.has RS232 interface, can be connected to computer or laboratory platform for data processing

19.provides mass data storage and printing functions, storing 1000 groups of data and supporting USB flash drives

 

LR-L100 Oil Particle Machine

Technical index

1.Light source: semiconductor laser

2.Particle size range: 0.8μm~600μm (according to 1~100um or 4~70um(c))

3.Detection channel: 8~64 channels Optional. Set any particle size

4.Sensitivity: 0.8μm (ISO4402) or 3μm (c) (GB/T18854, ISO11171)

5.Resolution: < 10% (GB/T18854, ISO11171)

6.Repeatability: RSD<2%

7.Sample viscosity: ≤650cSt (excessive viscosity can be detected with heating or dilution method)

8.Sampling volume: 0.2~1000mL, interval: 0.1mL

9.Sampling accuracy: better than ±0.5%

10.Sampling speed: 5 ~80mL/min

11.Pressure chamber: positive and negative pressure chamber device for sample degassing and high viscosity sample detection

12.Maximum vacuum of pressure chamber: -0.08Mpa

13.Maximum positive pressure of the pressure chamber: 0.8Mpa

14.Maximum particle concentration: 12000~40000 particles /mL

15.Temperature (optional) : Collection range: 1~100℃; Acquisition accuracy: 1℃

16.Water activity (optional) : Collection range: 1~100%RH; Acquisition accuracy: 5%RH

17.Water content (optional) : Collection range: 1~300ppm; Acquisition accuracy: 1ppm

18.Power supply: 110~245V AC, 50Hz, 70W

19.External dimension: 340mm╳410mm╳650mm

20.Net weight: 25kg






    has been added to your cart: