Description

LR-L300 Fluid Particle Counter

Main feature

1.The test principle of the light resistance (shading) method specified by the International Hydraulic Standards Committee

2.High precision laser sensor, wide test range, stable performance, low noise, high resolution

3.High precision bidirectional plunger metering pump sampling method, injection speed is adjustable, high sampling volume accuracy

4.pipe is made of 316L and PTFE material, which is resistant to corrosion and can meet the detection of various organic solvents and oil products

5.for laboratory or field measurement, optional decompression device for on-line high pressure measurement

6.An external pressure chamber can be connected to form positive/negative pressure for high viscosity sample detection and sample degassing

7.can use standard sampling bottles, sampling cups and other sampling containers, or directly connected to the hydraulic system online testing, to meet the testing requirements of different industries

8.Built-in multiple calibration curve, compatible with all common standards at home and abroad for calibration

9.Built-in GJB-420A, GJB-420B, NAS1638, ISO4406, SAE4059E and γ OCT17216 and other common standards, a test can give all the built-in standards under the data results, support custom standard testing, and can be set according to customer needs required standards

10.1000 particle size channels can be set up to facilitate particle size analysis

11.built-in data analysis system, can automatically determine the sample grade according to the standard, with automatic data processing, printing functions

12.Color touch screen operation, input in Chinese and English, with preset, input, modify, storage functions, easy to operate

13.has RS232 interface, can be connected to computer or laboratory platform for data processing, can also use USB for data storage

14.Built-in lithium battery, suitable for field work, no external power supply can be used

15.embedded design, high strength shell, easy to carry, suitable for all kinds of construction machinery

 

LR-L300 Fluid Particle Counter

Technical index

1.Light source: semiconductor laser

2.Particle size range: 0.8-600um (according to 1~100um or 4~70um(c) calibration)

3.Sensitivity: 0.1μm

4.Custom detection channel: 8+8 channels, any particle size

5.Sampling volume: 0.2-1000ml, interval: 0.1mL

6.Sampling accuracy: better than ±1%

7.Sampling speed: 5-80mL/min

8.Cleaning speed: 5-80mL/min

9.Counting accuracy: < ±5%

10.Resolution: ≤10%

11.Repeatability: RSD < 2%

12.limit coincidence error: 1200-40000 granules /mL

13.Off-line detection viscosity: ≤100cSt (if the viscosity is too large, an air pressure chamber can be selected)

14.Pressure range: low pressure 0-0.6Mpa, high pressure up to 30Mpa (optional pressure reducing valve)

15.Online detection interval: Set this parameter to any value

16.Sample temperature: 0℃~80℃

17.Operating temperature: -20℃~60℃

18.Storage temperature: -30℃~80℃

19.Power supply: AC100-240V, 50/60Hz

20.Battery capacity: 5200mAh

21.Battery running time: 6-8 hours

22.Dimensions: 410×320×165mm

23.Weight: 8.5kg






    has been added to your cart: