Description

LR-TMA105 TMA Analyzer

Overview:

LR-TMA105 thermomechanical analyzer is a thermomechanical analysis method (TMA) used to study the mechanical properties of polymer materials. It can determine the temperature and deformation curve of materials under the condition of constant temperature rise, so as to determine the thermal expansion coefficient, glass transition temperature Tg, flow temperature Tf and other characteristics of materials. It can be widely used in the research and production of polymer and its synthetic materials, drugs, ceramics and other materials.

 

LR-TMA105 TMA Analyzer

Meet the standard:

GB/T 36800.1-2018, GB/T36800.2-2018, GB 11998-1989, ISO 11359-2:1999

 

LR-TMA105 TMA Analyzer

Technical indicators:

1.Temperature of the experimental chamber

(1) Temperature range of the experimental chamber: room temperature ~ 500℃

(2) Temperature accuracy: ±0.5℃

(3) Heating rate: 0.5℃/ min ~ 10℃/ min 1,2,5 ℃/min commonly used

(4) Temperature control element: Pt100

(5) Furnace chamber temperature measuring unit: thermocouple

(6) Temperature resolution: 0.1℃

2, sample deformation displacement measurement

(1) Effective displacement measurement range: 0-2.5mm

(2) Displacement measurement resolution: 0.001mm

(3) Displacement measurement accuracy: ±0.005mm

(4) Displacement sensor measurement, RS232 serial port data output

3.Load part

(1) Load rod quality: 512g

(2) Weight: 400g (can be added according to the requirements of use)

4.Time display

Internal clock automatic calculation, time error: ±1s/h

5.Heating part

(1) Resistance wire heating controlled by solid state relay

(2) Heating voltage: AC220V, 50Hz

(3) Heating power: 800W

6.Test method

Stretch, bend, compress, pin,

7.Tensile sample:

Maximum clamping thickness: < 3mm

8.Bending sample:

(1) Bending head radius: R3.0± 0.10mm

(2) Bending fulcrum radius: R3.0± 0.10mm

(3) Bending support spacing: 15.0mm

9.Compression test:

(1) Head area: 12.56mm2

(2) Indenter diameter: Φ4.0± 0.05mm

10.Needle insertion test

(1) Needle area: 1.00 mm2

(2) Needle diameter: Φ1.13± 0.05mm

11.Total power supply: 1000W

12, Power supply: 220V±10%, 50Hz

13, the instrument size: 350mm× 300mm × 600mm

14, instrument weight: about 15kg






    has been added to your cart: